Abstract
Background
Internal fixation is one of the main options for treating displaced intracapsular
hip fractures. However, controversy remains over which osteosynthesis is the best
choice. Using a simulated displaced intracapsular hip fracture model, we compared
the mechanical stability of three types of osteosynthesis: the sliding hip screw (SHS),
three cannulated screws and the Targon® FN. We also assessed whether bone mineral density (BMD) influenced the stability
of the fixation.
Methods
Unstable/displaced intracapsular hip fractures were induced in a total of 12 pairs
of fresh-frozen cadaver femora. Each fracture was fixed at random on the left or right
side with an SHS or three cannulated screws (six bone pairs; study 1), or with an
SHS or the Targon® FN implant (six bone pairs; study 2). All femoral heads were exposed to cyclic combined
axial and torque loads until failure. The failure mechanism, the maximal load-to-failure
and the dual-energy X-ray absorptiometry (DEXA) values of the femoral heads were determined
and their relationships were analysed.
Results
There was no significant difference in the maximal load-to-failure between the SHS
and the three cannulated screws. The load-to-failure was significantly higher for
the Targon® FN than for the SHS. There was a high correlation between the bone mineral densities
(BMDs) of the femoral heads and maximal load-to-failure in the Targon® FN group only.
Interpretation
Basing the implant choice on preimplantation BMD measurements does not ensure the
best biomechanical outcome. We found that the combination of a fixed-angle device
and multiple sliding neck screws (Targon® FN) enhances the mechanical strength of reconstructions in unstable/displaced intracapsular
hip fractures.
Keywords
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to InjuryAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Femoral neck fracture fixation: a biomechanical study of two cannulated screw placement techniques.Orthopedics. 1998; 21: 1173-1176
- Biomechanical analysis of the percutaneous compression plate and sliding hip screw in intracapsular hip fractures: experimental assessment using synthetic and cadaver bones.Injury. 2006; 37 ([Epub 2006]): 979-983
- The displaced femoral neck fracture. Internal fixation versus bipolar endoprosthesis. Results of a prospective, randomized comparison.Clin Orthop. 1988; 230: 127-140
- Statistical power analysis for the behavioural sciences.2nd ed. Lawrence Erlbaum, New Jersey1988
- Hip fractures in the elderly: a world-wide projection.Osteoporos Int. 1992; 2: 285-289
- Biomechanical comparison of internal fixation techniques for the treatment of unstable basicervical femoral neck fractures.J Orthop Trauma. 1997; 11: 337-343
- Fixation of 220 femoral neck fractures. A prospective comparison of the Rydell nail and the LIH hook pins.Acta Orthop Scand. 1990; 61: 154-157
- Early loss of fixation of femoral neck fractures. Comparison of three devices in 244 cases.Acta Orthop Scand. 1989; 60: 69-72
- Outcome after displaced fracture of the femoral neck: a meta-analysis of one hundred and six published reports.J Bone Joint Surg. 1994; 76-A: 15-25
- Fixation of displaced femoral neck fractures. A comparison between sliding screw plate and four cancellous bone screws.Acta Orthop Scand. 1987; 58: 212-216
- Two or three screws for fixation of femoral neck fractures?.Am J Orthop. 2003; 32: 438-442
- Treatment of femoral neck fractures with a sliding compression screw and two Knowles pins.Clin Orthop. 1984; 190: 158-162
- Hip fracture management.Blackwell Scientific Publications, Oxford1993
- New developments in hip fracture treatment.Trauma. 2003; 5: 43-49
- Internal fixation implants for intracapsular proximal femoral fractures in adults.Cochrane Database Syst Rev. 2008;
- Subchondral screw fixation for femoral neck fractures.J Bone Joint Surg. 1989; 71: 178-180
- Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study.Injury. 2004; 35: 136-141
- Bone mineral content and fixation strength of femoral neck fractures. A cadaver study.Acta Orthop Scand. 1994; 65: 161-165
- Biomechanical factors and failure of transcervical hip fracture repair.Injury. 2001; 32: 223-228
- The effect of screw insertion site and unused drill holes on stability and mode of failure after fixation of basicervical femoral neck fracture.Crit Rev Biomed Eng. 2000; 28: 11-16
Article info
Publication history
Accepted:
August 18,
2010
Identification
Copyright
© 2010 Elsevier Ltd. Published by Elsevier Inc. All rights reserved.