Biomechanical analysis of the sliding hip screw, cannulated screws and Targon® FN in intracapsular hip fractures in cadaver femora

      Abstract

      Background

      Internal fixation is one of the main options for treating displaced intracapsular hip fractures. However, controversy remains over which osteosynthesis is the best choice. Using a simulated displaced intracapsular hip fracture model, we compared the mechanical stability of three types of osteosynthesis: the sliding hip screw (SHS), three cannulated screws and the Targon® FN. We also assessed whether bone mineral density (BMD) influenced the stability of the fixation.

      Methods

      Unstable/displaced intracapsular hip fractures were induced in a total of 12 pairs of fresh-frozen cadaver femora. Each fracture was fixed at random on the left or right side with an SHS or three cannulated screws (six bone pairs; study 1), or with an SHS or the Targon® FN implant (six bone pairs; study 2). All femoral heads were exposed to cyclic combined axial and torque loads until failure. The failure mechanism, the maximal load-to-failure and the dual-energy X-ray absorptiometry (DEXA) values of the femoral heads were determined and their relationships were analysed.

      Results

      There was no significant difference in the maximal load-to-failure between the SHS and the three cannulated screws. The load-to-failure was significantly higher for the Targon® FN than for the SHS. There was a high correlation between the bone mineral densities (BMDs) of the femoral heads and maximal load-to-failure in the Targon® FN group only.

      Interpretation

      Basing the implant choice on preimplantation BMD measurements does not ensure the best biomechanical outcome. We found that the combination of a fixed-angle device and multiple sliding neck screws (Targon® FN) enhances the mechanical strength of reconstructions in unstable/displaced intracapsular hip fractures.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Injury
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Booth K.C.
        • Donaldson T.K.
        • Dai Q.G.
        Femoral neck fracture fixation: a biomechanical study of two cannulated screw placement techniques.
        Orthopedics. 1998; 21: 1173-1176
        • Brandt E.
        • Verdonschot N.
        • van Vugt A.
        • van Kampen A.
        Biomechanical analysis of the percutaneous compression plate and sliding hip screw in intracapsular hip fractures: experimental assessment using synthetic and cadaver bones.
        Injury. 2006; 37 ([Epub 2006]): 979-983
        • Bray T.J.
        • Smith-Hoefe E.
        • Hooper A.
        • Timmerman L.
        The displaced femoral neck fracture. Internal fixation versus bipolar endoprosthesis. Results of a prospective, randomized comparison.
        Clin Orthop. 1988; 230: 127-140
        • Cohen J.
        Statistical power analysis for the behavioural sciences.
        2nd ed. Lawrence Erlbaum, New Jersey1988
        • Cooper C.
        • Campion G.
        • Melton L.J.
        Hip fractures in the elderly: a world-wide projection.
        Osteoporos Int. 1992; 2: 285-289
        • Deneka D.A.
        • Simonian P.T.
        • Stankewich C.J.
        • et al.
        Biomechanical comparison of internal fixation techniques for the treatment of unstable basicervical femoral neck fractures.
        J Orthop Trauma. 1997; 11: 337-343
        • Holmsberg S.
        • Mattson P.
        • Dahlborn M.
        • Ersmark H.
        Fixation of 220 femoral neck fractures. A prospective comparison of the Rydell nail and the LIH hook pins.
        Acta Orthop Scand. 1990; 61: 154-157
        • Husby T.
        • Alho A.
        • Nordsletten L.
        • Bugge W.
        Early loss of fixation of femoral neck fractures. Comparison of three devices in 244 cases.
        Acta Orthop Scand. 1989; 60: 69-72
        • Lu-Yao G.L.
        • Keller R.B.
        • Littenberg B.
        • Wennberg J.E.
        Outcome after displaced fracture of the femoral neck: a meta-analysis of one hundred and six published reports.
        J Bone Joint Surg. 1994; 76-A: 15-25
        • Madsen F.
        • Linde F.
        • Anderson E.
        • et al.
        Fixation of displaced femoral neck fractures. A comparison between sliding screw plate and four cancellous bone screws.
        Acta Orthop Scand. 1987; 58: 212-216
        • Maurer S.G.
        • Wright K.E.
        • Kummer F.J.
        • et al.
        Two or three screws for fixation of femoral neck fractures?.
        Am J Orthop. 2003; 32: 438-442
        • Ort P.J.
        • LaMont J.
        Treatment of femoral neck fractures with a sliding compression screw and two Knowles pins.
        Clin Orthop. 1984; 190: 158-162
        • Parker M.J.
        • Pryor G.A.
        Hip fracture management.
        Blackwell Scientific Publications, Oxford1993
        • Parker M.J.
        New developments in hip fracture treatment.
        Trauma. 2003; 5: 43-49
        • Parker M.J.
        • Stockton G.
        • Gurusamy K.
        Internal fixation implants for intracapsular proximal femoral fractures in adults.
        Cochrane Database Syst Rev. 2008;
        • Rehnberg L.
        • Olerud C.
        Subchondral screw fixation for femoral neck fractures.
        J Bone Joint Surg. 1989; 71: 178-180
        • Selvan V.T.
        • Oakley M.J.
        • Rangan A.
        • Al-Lami M.K.
        Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study.
        Injury. 2004; 35: 136-141
        • Sjöstedt A.
        • Zetterberg C.
        • Hansson T.
        • et al.
        Bone mineral content and fixation strength of femoral neck fractures. A cadaver study.
        Acta Orthop Scand. 1994; 65: 161-165
        • Spangler L.
        • Cummings P.
        • Tencer A.F.
        • et al.
        Biomechanical factors and failure of transcervical hip fracture repair.
        Injury. 2001; 32: 223-228
        • Stafford P.
        • Goulet R.
        • Norris B.
        The effect of screw insertion site and unused drill holes on stability and mode of failure after fixation of basicervical femoral neck fracture.
        Crit Rev Biomed Eng. 2000; 28: 11-16

      Linked Article