Advertisement
Paper| Volume 26, ISSUE 7, P483-485, September 1995

Stability of reamed and unreamed intramedullary tibial nails: a biomechanical study

      This paper is only available as a PDF. To read, Please Download here.

      Abstract

      We examined the correlation of bone mineral density and bone strength in the cadaver tibia, and looked in vitro at the relative stability of tibial fractures fixed with either reamed or unreamed libial nails. Bonemineral density correlated well with bone strength (r = 0.946), but paired tibias did not correlate closely. The unreamed nail-bone construct was less stable than the reamed construct in each pair tested (P<0.05), and a comparison of all bones showed it to be less stable at all levels of our testing regimen, including failure (P<0.01). Bending and breakage was seen in four of the smaller unreamed interlocking screws (4 mm).

      References

        • Müller ME
        • Allgöwer M
        • Schneider R
        • et al.
        2nd Ed. Manual of Internal Fixation. Springer, New York1979
        • Haas N
        • Gotzen L Die
        Tibiaschaftfrakiur des Erwaschsenen.
        in: Schmit-Neuerburg KP Stürmer KP Plattenosteosynthese. Springer, Berlin, Heidelberg, New York1989
        • Syszkowitz R
        • Reschauer
        • Seggl W
        Gefahren der Plattenosteosynthese und Möglichkeiten des Fixateur exteme in der Frakturvesorgung.
        Heft Unfallheilkunde. 1981; 153 (or 1989): 179
        • Gustilo B
        • Anderson JP
        Prevention of infection on the treatment of one thousand and twenty five open fractures of long bones.
        J Bone Joint Surg [Am]. 1976; 58A: 453
        • Caudle RJ
        • Stern PJ
        Severe open fractures of the tibia.
        J Bone Joint Surg [Am]. 1987; 69A: 801
        • Holbrook JL
        • Swiontkowski MF
        • Sanders R
        Treatment of open fractures of the tibial shaft: Ender nailing versus external fixation. A randomized, prospective comparison.
        J Bone Joint Surg [Am]. 1989; 71A: 1231
        • Wiss DA
        Flexible medullary nailing of acute tibial shaft fractures.
        Clin Orthop. 1986; 212: 122
        • Krettek C
        • Haas N
        • Tscherne H
        Behandlungsergebnisse von 202 frischen Unterschenkelschaftfrakturen, versorgt mit einem unilateralen Fixateur externe (Monofixateur).
        Unfallheikunde. 1988; 92: 440
        • Etter C
        • Burri C
        • Claes L
        • et al.
        Treatment by external fixation of open fractures associated with severe soft tissue damage of the leg.
        Clin Orthop. 1983; 178: 80
        • Maatz R
        Zur infekthäufigkeit nach gedeckter oder offner Nagelung geschlossener Frakturen.
        Aktuel Traumatol. 1983; 13: 175
        • Küntschner G
        3rd Ed. Praxis der Marknagelung. Schattauer, Stuttgart1962
        • Harvey FJ
        • Hodkinson AH
        • Harvey PM
        Intramedullary nailing in the treatment of open fractures of the tibia and fibula.
        J Bone Joint Surg [Am]. 1975; 57A: 909
        • Povacz F
        Verbrennungsschaden an der Tibiadiaphyse nach Marknagelung mit aufbohren.
        Unfallheilkunde. 1979; 82: 126
        • Velasco A
        • Whiteside Jr, TE
        • Fleming LL
        Open fractures of the tibia treated with Lottes nail.
        J Bone Joint Surg [Am]. 1983; 65A: 879
        • Chapman MW
        The role of intramedullary nailing in open fracture.
        Clin Orthop. 1986; 212: 26
        • Kohlman H
        • Vescei V
        • Rabitsch K
        • et al.
        Zur inditaktion der Verriegelungsnagelung bei offnen Frakturen.
        Aktuel Traumatol. 1988; 18: 59
        • Brumback RJ
        • Ellison Jr, PS
        • Poka A
        • et al.
        Intramedullary nailing of open fractures of the femoral shaft.
        J Bone Joint Surg [Am]. 1989; 71A: 1324
        • Lhowe DW
        • Hansen ST
        Immediate nailing of open fractures of the femoral shaft.
        J Bone Joint Surg [Am]. 1988; 70A: 812